Emerson MICRO MOTION 1700 User Manual

Page 74

Advertising
background image

3.

Screw the nipple into the conduit opening for the 9-wire cable. Tighten it to one turn
past hand-tight.

4.

Slide the compression ring, compression nut, and sealing nut onto the cable. Make
sure the compression ring is oriented so the taper will mate properly with the
tapered end of the nipple.

5.

Pass the cable end through the nipple so the braided shield slides over the tapered
end of the nipple.

6.

Slide the compression ring over the braided shield.

7.

Screw the compression nut onto the nipple. Tighten the sealing nut and
compression nut by hand to ensure that the compression ring traps the braided
shield.

8.

Use a 25-mm (1-inch) wrench to tighten the sealing nut and compression nut to
20–25 foot-pounds (27–34 N-m) of torque.

Cross-section of assembled cable gland with cable

Figure 5-21:

A

B

C

D

E

F

G

A

A. Cable
B. Sealing nut
C. Seal
D. Compression nut
E. Braided shield
F. Brass compression ring
G. Nipple

9.

Remove the junction box cover and remote core processor end-cap.

10.

At both the sensor and remote core processor, connect the cable according to the
following procedure:

a. Insert the stripped end of each wire into the corresponding terminal at the

sensor and remote core processor ends, matching by color. No bare wires should
remain exposed.

Mounting and sensor wiring for remote core processor with remote sensor installations