Specifications, Electrical connections, Chapter 10 – Emerson MICRO MOTION 1700 User Manual

Page 113: 10 specifications, 1 electrical connections

Advertising
background image

10

Specifications

Topics covered in this chapter:

Electrical connections

Input/output signals

Local display

Environmental limits

Physical specifications

10.1

Electrical connections

Electrical connections

Table 10-1:

Type

Description

Input/output connections

Three pairs of wiring terminals for transmitter outputs. Screw
connectors accept one or two solid conductors, 14 to 12 AWG
(2.5 to 4.0 mm

2

); or one or two stranded conductors, 22 to

14 AWG (0.34 to 2.5 mm

2

).

Power connections

One pair of wiring terminals accepts AC or DC power.
One internal ground lug for power-supply ground wiring.
Screw connectors accept one or two solid conductors, 14 to
12 AWG (2.5 to 4.0 mm

2

); or one or two stranded conductors, 22

to 14 AWG (0.34 to 2.5 mm

2

).

Digital communications main-
tenance connections

Two clips for temporary connection to the service port.

Core processor connection

The transmitter has two pairs of terminals for the 4-wire connec-
tion to the core processor:

One pair is used for the RS-485 connection to the core pro-
cessor

One pair is used to supply power to the core processor

Plug terminals accept solid or stranded conductors, 24 to
12 AWG (0.40 to 3.5 mm

2

).

Power

Self-switching AC/DC input, automatically recognizes supply
voltage

85 to 265 VAC, 50/60 Hz, 6 watts typical, 11 watts maximum

18 to 100 VDC, 6 watts typical, 11 watts maximum

Complies with low voltage directive 2006/95/EC per EN
61010-1 (IEC 61010-1) with amendment 2, and Installation
(Overvoltage) Category II, Pollution Degree 2

Specifications