HART/analog single loop wiring

Note

For HART communications:

600 Ω maximum loop resistance

250 Ω minimum loop resistance

HART/analog single loop wiring

Figure 7-2:

A

B

A.

820 Ω maximum loop resistance

B.

HART-compatible host or controller

I/O wiring for Model 1700 and Model 2700 transmitters with analog outputs