Preparing jacketed cable

Figure 5-11:

1. Trim 4 inches (100 mm) of cable jacket.
2. Remove the clear wrap and filler material.
3. Remove the foil that is around the insulated wires

and separate them.

4. Identify the drain wires in the cable and bring them

together. Fan the other wires to the outside of the
cable. Twist the drain wires together.

5. Slide the 3-inch (75 mm) heat-shrink tubing over

the drain wires. Push the tubing as close as possible
to the cable jacket.

6. Slide the 1 ½ inch (40 mm) long heat-shrink tubing

over the cable jacket. The tubing should
completely cover all portions of the drain wires that
remain exposed next to the cable jacket.

6. Without burning the cable, apply heat to shrink all

tubing. Recommended temperature is 250 °F (121
°C).

1. Trim 4 ½ inches (115 mm) of cable jacket.
2. Remove the clear wrap and filler material.
3. Remove the foil that is around the insulated wires

and separate them.

4. Identify the drain wires in the cable. Clip off each

drain wire as close as possible to the cable jacket.

7. Allow the cable to cool, then strip ¼ inch (5 mm) of

insulation from each wire.

5. Slide the 1 ½ inch (40 mm) heat-shrink tubing over

the wires and cable jacket. The tubing should
completely cover the clipped ends of the drain wires.

Trim cable jacket

Drain wires clipped

Heat-shrink

tubing

Trim cable jacket

Heat-shrink tubing over

cable jacket

7. Without burning the cable, apply heat to shrink all

tubing. Recommended temperature is 250 °F (121
°C).

8. Allow the cable to cool, then strip ¼ inch (5 mm) of

insulation from each wire.

Heat-shrink tubing over drain

wires

Prepare jacketed

cable at the

transmitter end

Prepare jacketed

cable at the sensor

end

Mounting and sensor wiring for remote core processor with remote sensor installations