Emerson MICRO MOTION 1700 User Manual

Page 109

Advertising
background image

Recommended pull-up resistor versus supply voltage

Figure 9-15:

3600

3200

2800

2400

2000

1600

1200

800

0

5

10

15

20

25

30

Supply voltage (Volts)

External pull-up resistor range (Ohms)

4000

4400

9.4.3

Internally powered discrete output wiring on Channel C

Internally powered discrete output wiring on Channel C

Figure 9-16:

A

A.

Total load

I/O wiring for Model 2700 transmitters with configurable input/outputs