Preparing shielded or armored cable

Figure 4-4:

1. Without cutting the shield, strip 9 inches (225 mm) of

cable jacket.

2. Strip 8 ½ inches (215 mm) of braided shield, so ½

inch (10 mm) of shield remains exposed.

3. Remove the foil shield that is between the braided

shield and inner jacket.

4. Strip 4 inches (100 mm) of inner jacket.

5. Remove the clear wrap and filler material.
6. Remove the foil that is around the insulated wires and

separate them.

7. Identify the drain wires in the cable and bring them

together. Fan the other wires to the outside of the
cable. Twist the drain wires together.

8. Slide the 3-inch (75 mm) long heat-shrink tubing over

the drain wires. Push the tubing as close as possible to
the inner jacket.

9. Slide the 1 ½ inch (40 mm) long heat-shrink tubing

over the cable jacket. The tubing should completely
cover all portions of the drain wires that remain
exposed next to the cable jacket.

9. Without burning the cable, apply heat to shrink all

tubing. Recommended temperature is 250 °F (121
°C).

1. Without cutting the shield, strip 7 inches (175 mm)

of outer jacket.

2. Strip 6 ½ inches (165 mm) of braided shield, so ½

inch (10 mm) of shield remains exposed.

3. Remove the foil shield that is between the braided

shield and inner jacket.

4. Strip 4 ½ inches (115 mm) of inner jacket.

5. Remove the clear wrap and filler material.
6. Remove the foil that is around the insulated wires

and separate them.

7. Identify the drain wires in the cable. Clip each drain

wire as close as possible to the cable jacket.

10. Allow the cable to cool, then strip ¼ inch (5 mm) of

insulation from each wire.

8. Slide the 1 ½ inch (40 mm) long heat-shrink tubing

over the cable jacket. The tubing should completely
cover the clipped ends of the drain wires.

Drain wires clipped

Heat-shrink tubing

Heat-shrink tubing over cable

jacket

Trim inner jacket

Trim braided shield

Trim outer jacket

Trim inner jacket

Trim braided shield

Trim outer jacket

Prepare shielded or

armored cable at the

sensor end

Prepare shielded or

armored cable at the

transmitter end

Heat-shrink tubing over drain wires

10. Without burning the cable, apply heat to shrink all

tubing. Recommended temperature is 250 °F (121
°C).

11. Allow the cable to cool, then strip ¼ inch (5 mm) of

insulation from each wire.

4.2.1

Micro Motion 9-wire cable types and usage

Mounting and sensor wiring for 9-wire remote installations