Safe area mA output load resistance values

Figure 8-4:

1000

900

800

700

600

500

400

300

200

100

0

12

14

16

18

20

22

24

26

28

30

OPERATING REGION

Supply voltage VDC (Volts)

External resistor R

load

(Ohms)

R

max

= (V

supply

– 12)/0.023

Min. 250W and 17.5V required for HART communication

8.3

Safe area HART multidrop wiring

Tip

For optimum HART communication, single-point ground the output loop to an instrument-grade
ground.