Safe area HART multidrop wiring

Figure 8-5:

A

B

C

D

E

F

A.

250–600 Ω resistance

B.

HART-compatible host or controller

C.

HART-compatible transmitters

D.

Model 1700 or Model 2700 transmitter with intrinsically safe outputs

E.

SMART FAMILY transmitter

F.

24 VDC loop power supply required for HART 4–20 mA passive transmitters

8.4

Safe area frequency output/discrete output

wiring

Safe area frequency output/discrete output wiring

Figure 8-6:

A

B

c

A.

VDC

B.

Counter

C.

R

load

I/O wiring for Model 1700 and Model 2700 transmitters with intrinsically safe outputs