Hazardous area wiring, 5 hazardous area wiring – Emerson MICRO MOTION 1700 User Manual

Page 91

Advertising
background image

Safe area frequency output/discrete output load resistance values

Figure 8-7:

10000

9000

8000

7000

6000

5000

4000

3000

2000

1000

0

5

7

9

11

13

15

17

19

21

23

OPERATING REGION

Supply voltage VDC (Volts)

External pull-up resistor R

load

range

(Ohms)

R

max

= (V

supply

– 4)/0.003

Rmin = (Vsupply – 25)/0.006

Min. 100 W for supply voltage less than 25.6 Volts

25

27

29

8.5

Hazardous area wiring

Information provided about I.S. barriers is intended as an overview. Application-specific or
product-specific questions should be addressed to the barrier manufacturer or
Micro Motion.

DANGER!

Hazardous voltage can cause severe injury or death. Shut off the power before wiring
transmitter outputs.

DANGER!

Improper wiring in a hazardous environment can cause an explosion. Install the transmitter
only in an area that complies with the hazardous classification tag on the transmitter.

Safety parameters

Table 8-1:

Parameter

4–20 mA

Frequency/discrete

Voltage (U

i

)

30 V

30 V

Current (I

i

)

300 mA

100 mA

Power (P

i

)

1.0 W

0.75 W

I/O wiring for Model 1700 and Model 2700 transmitters with intrinsically safe outputs

Installation Manual

87

Advertising
This manual is related to the following products: