User calibration, P189), 1-1 user calibration – Sekonic Speedmaster L-858D-U Light Meter User Manual

Page 203

Advertising
background image

189

7. Hardware Setting

Operation

<When you know the compensation value in advance>

1.

Touch the [User Calibration] Button.

The User Calibration Screen is displayed.

User Calibration Screen

[Close] Button

Calibration Value

Adjustment Button

(+)

Adjustment Button

(-)

2.

Adjust the calibration value.

Use the Adjustment Button (

) or (

) to adjust the calibration value in

increments of +/-0.1 EV.

3.

Touch [Close] Button.

Finish the setting, and return to the Hardware Setting Screen.

7-1-1

User Calibration

The meter is calibrated to Sekonic standards. However, if necessary, you can

change the measurement standard using the User Calibration Function. The

compensation value can be set +/- 1.0 EV in 0.1 EV step increments.

If you know the compensation value in advance, you can directly enter the value.

Also, you can adjust the meter based on a measured value obtained using another

light meter.