Sekonic Speedmaster L-858D-U Light Meter User Manual

Page 204

Advertising
background image

190

7. Hardware Setting

<When adjusting the meter based on a measured value

obtained using another light meter>

1.

Touch the [User Calibration] Button.

The User Calibration Screen is displayed.

User Calibration Screen

[Close] Button

Calibration

Value

Adjustment Button

(+)

Adjustment Button

(-)

2.

Press the Measuring Button

6

.

The User Calibration Screen allows you to make measurements, thus you can equalize

measured values obtained using another light meter under the same light source.

3.

Adjust the calibration value.

Touch the Adjustment Button (

) or (

) to adjust the calibration value in

increments of +/-0.1 EV.

4.

Touch [Close] Button.

Finish the setting, and return to the Hardware Setting Screen.

NOTICE

User calibration carried out in Hardware Setting is not displayed on the

status bar.

● Calibration of the measured value must be performed based on a sufficient

number of test shooting results.

Note that the individual compensation is possible in the incident light

system and reflected light system while uniform compensation is applied

in Ambient Mode and Flash Mode.