Calibration control register, Read, Write – HP E1418A User Manual

Page 130

Advertising
background image

Calibration Control

Register

The Calibration Control Register sets and reports calibration conditions.

Address

Base + 08

16

Bit

15

14

13

12

11

10

09

08

07

06

05

04

03

02

01

00

READ

Cal

Bus En

A/D
Bsy

Cal Mode Cal Mux

En

Cal Mux Chan

Address

A/D Reading Value

WRITE

Cal

Bus En

A/D

Trg

Cal Mode

Sel

Cal Mux

En

Cal Mux Chan

Address

A/D

Scale

Unused

READ

Cal Bus En: A one (1) in this field indicates the cal bus is connected. A
zero (0) in this field indicates the cal bus is not connected. This bit reflects
the state of the last write to Cal Bus En field.

A/D Bsy: A one (1) in this field indicates the A/D is taking a reading. A
zero (0) in this field indicates a new reading is available.

Cal Mode: A one (1) in this field indicates the calibration system is set to
calibrate voltage. A zero (0) in this field indicates the calibration system is
set to calibrate current. This bit reflects the state of the last write to Cal
Mode Sel
field.

Cal Mux En: A one (1) in this field indicates the Cal Mux is enabled. A
zero (0) in this field indicates the Cal Mux is disabled (high-impedance).
This bit reflects the state of the last write to Cal Mux En field.

Cal Mux Chan Address: This field returns the channel selected for
calibration. This bit reflects the state of the last write to Cal Mux Chan
Address
field. The channel address, expressed in hexadecimal, has values
from 0

16

through F

16

corresponding to channels 1 through 16, respectively.

A/D Reading Value: This field returns the latest reading taken by the A/D.
When A/D Scale is set to 0, the actual voltage represented by the reading
value is given as:

Voltage

=

16.8 V

33.6 V

255 steps

AD Reading Value

When A/D Scale is set to 1, the actual voltage represented by the reading
value is given as:

Voltage

=

16.8 V

33.6 V

255 steps

AD Reading Value

13.7

WRITE

Cal Bus En: Writing a one (1) in this field connects the cal bus. Writing a
zero (0) in this field disconnects the cal bus.

A/D Trg: Writing a one (1) in this field begins an A/D reading. Writing a zero
(0) in this field during a reading in progress will corrupt the reading. Do not
combine this bit setting with other bits in this register (write only this bit).

Appendix B

HP E1418A Register-Based Programming 129

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Advertising