STATus:OPERation:CONDition?

STATus:OPERation:CONDition?

returns the contents of the Operation Status

Condition Register. Reading the register does not affect its contents.

Parameters

Returns: 16-bit signed integer (int16).

Comments

Bit 0 of this register (decimal 1) is the Calibrating bit. This bit is set when:

the

CALibration:CONFigure:RESistance?

,

CALibration

n

:CONFigure:VOLTage?

,

or

CALibration

n

:CONFigure:CURRent?

command has been received and,

measurements with the corresponding

CALibration:VALue:RESistance

,

CALibration

n

:VALue:VOLTage?

, or

CALibration

n

:VALue:CURRent?

commands

are not complete.

*RST Condition: clears all conditions.

*CLS

and

STATus:PRESet

do not affect bits in the Status Operation Condition

Register.

STATus:OPERation:ENABle

STATus:OPERation:ENABle <

mask

>

specifies which bits of the Operational

Status Event Register are OR’d together to create the Operational Status bit (bit 7)
in the Summary Register of the Status Byte.

Parameters

Parameter

Name

Parameter

Type

Range of

Values

Default

<

mask

>

numeric (int16) or

non-decimal numeric

-32768 to +32767 or #H0000 to #HFFFF

none

Comments

Only bit 0 (decimal 0) is used or defined in the Operation Register. Bit 0 is
defined as the Calibrating bit. A 1 in this bit indicates the module is currently in
the calibration mode.

The non-decimal numeric forms are the #H, #Q, or #B formats specified by
IEEE-488.2.

*RST

clears all conditions.

*CLS

and

STATus:PRESet

do not affect bits in the Operation Status Condition

Register.

STATus

96 HP E1418A SCPI Command Reference