Calibration:configure:resistance, Calibration:reset, Calibration – HP E1418A User Manual

Page 69

Advertising
background image

CALibration:CONFigure:RESistance?

CALibration:CONFigure:RESistance?

configures the module to perform

calibration of the internal calibration resistor and returns a

0

or

1

indicating the

readiness of the module.

Parameters

Returns: signed 16-bit integer (int16).

1

= ready,

0

= not ready.

Comments

The internal calibration resistor should be calibrated before calibrating the
channels in current mode.

This query configures the module for a 4-wire resistance measurement at the CAL
output terminals when the calibration point is set to

CALB

(following a

*RST

or

DIAGnostic:CALibration:OUTPut[:MODE] CALBus

command).

This query may also be used to set an external calibration resistor value when
calibrating at each channel output.

This query is used together with the

CALibration:VALue:RESistance

command to

calibrate the current calibration resistor.

Refer to Appendix D, “ Voltage/Current Output Adjustment” , starting on page 154,
for additional information about this command.

Related Commands:

CALibration:VALue:RESistance

CALibration:RESet

CALibration:RESet

reloads, for module use, the calibration constants stored in

non-volatile memory.

Parameters

None.

Comments

This command, with respect to the calibration constants, performs the same action
as power-on.

At power-on, or following this command, calibration constants stored in
non-volatile memory are loaded into the module where they are used during
output operations.

If the trigger system is waiting for a trigger when this command is received,
an error is generated.

This command opens all output relays.

Related Commands:

CALibration:STORe

CALibration

68 HP E1418A SCPI Command Reference

Chapter 3

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Advertising