Source n :voltage[:level]:triggered[:amplitude, Source – HP E1418A User Manual

Page 95

Advertising
background image

SOURce

n

:VOLTage[:LEVel]:TRIGgered[:AMPLitude]?

SOURce

n

:VOLTage[:LEVel]:TRIGgered][:AMPLitude]? [<MIN | MAX | DEF>

queries the latest value written by either

SOURce

n

:VOLTage[:LEVel]:TRIGgered[:AMPLitude]

or

SOURce

n

:VOLTage[:LEVel][:IMMediate][:AMPLitude]

.

Parameters

Parameter

Name

Parameter

Type

Range of

Values

Default

SOURce

n

keyword

1 to 8 or 1 to 16

1

<MIN | MAX | DEF>

discrete (optional)

MIN | MAX | DEF

none

Returns: 32-bit floating point number (float 32).

Comments

In the keyword,

SOURce

n, the valid range of n is:

1 to 8 for

E1418A, E1418A Option 002

1 to 16 for

E1418A Option 001, E1418A Option 003.

Values of n outside these ranges will generate an error.

If no parameter is given, this query returns the last value set on channel n by the

SOURce

n

:VOLTage[:LEVel][:IMMediate][:AMPLitude]

or

SOURce

n

:VOLTage[:LEVel]:TRIGgered[:AMPLitude]

command. The value

returned is the value last programmed on that channel, not an actual measurement
of the hardware output value.

If a parameter is given,

MIN

returns -16.000.

MAX

returns +16.000.

DEF

returns 0.000.

This command generates an error if channel n is in the current output mode.

Related Commands:

SOURce

n

:VOLTage[:LEVel]:TRIGgered[:AMPLitude]

,

SOURce

n

:FUNCtion:MODE VOLTage

SOURce

94 HP E1418A SCPI Command Reference

Chapter 3

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Advertising