err=viScanf(dac, “%d”, &result);

/* Enter integer returned */
/*, returns a 1 when ready to calibrate */

if(err < VI_SUCCESS) err_handler(dac, err);
}

result = 1;
while (result !=0 && result !=2)

{
meas_volt=get_voltage ();

/* Get the voltage measurement */

err=viPrintf(dac, “CAL%d:VAL:CURR? %f\n”,i,meas_volt);
if(err < VI_SUCCESS) err_handler(dac, err);

err=viScanf(dac, “%d”, &result);

/* Enter integer returned */
/* returns a 0 when calibration complete */

if(err < VI_SUCCESS) err_handler(dac, err);

if (result == 5)

/* Calibration aborted */

{

printf (“Calibration for channel %d aborted\n\n”,i);

result = 0;

}

}

}

/*** STORE THE NEW CALIBRATION CONSTANTS ***/

err=viPrintf(dac, “CAL:STOR;*OPC?\n”);

/* Store the new cal constants */

if(err < VI_SUCCESS) err_handler(dac, err);
err=(viScanf(dac,"%d",&result);
if(err < VI_SUCCESS) err_handler(dac, err);
sys_err(dac);

}

Appendix D

Voltage/Current Output Adjustment 164

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Advertising