Table C-2. Error Messages

Code

Error Messages

Potential Cause(s)

–141 Invalid character data

An invalid SCPI keyword was used.

–211 Trigger ignored

A trigger was received and no channels were in the wait for
trigger state.

–213 Init ignored

An INITiate was received and one or more channels were

already in the wait for trigger state.

–221 Setting Conflict

A command was received for a mode not set. For example, a
SOUR

n

:VOLT command was received and the channel is set to

CURR mode.

–222 Data out of range

<

n

> in SCPI keywor

d

i

s not in the range of 1 to 16.

<

value

> is out of range.

<

mask

> is out of range.

<

number

> is out of range.

–224 Illegal Parameter Value

Value received is not in the legal range for the parameter.

–240 Hardware Error

A hardware failure has been detected. The module may require
replacement or repair.

+2001 Invalid Channel Number

<

n

> in SCPI keyword is not valid for the module configuration.

For example <

n

> is set to 11 on an 8-channel configuration.

+3000 Trigger too fast

The trigger was received before the module completed the last
command.

+3002 Self Test Failed

A hardware failure was detected during power up self-test.
Additional self-test information is given in Appendix E,
beginning on page 166.

+3004 Multiple attempts to erase

non-volatile Memory failed

Attempts to erase non-volatile memory using the CAL:STOR
command failed. May indicate a hardware failure or the CAL
Store Enable Jumper is not in the correct position.

+3005 Multiple attempts to program

non-volatile Memory failed

Attempts to write to non-volatile memory using the CAL:STOR
command failed. May indicate a hardware failure or the CAL
Store Enable Jumper is not in the correct position.

+3006 CAL Store Enable Jumper not set

properly

A CAL:STOR command was received and the CAL Store
Enable Jumper is not set to the CAL position.

+3007 Checksum error on non-volatile

Memory

Calibration constants in non-volatile memory are corrupted.
Recalibration is needed.