Calibration resistor value registers, Voltage calibration status register – HP E1418A User Manual

Page 145

Advertising
background image

Calibration

Resistor Value

Registers

The Calibration Resistor Value Registers store the value of the resistor used
to calibrate current output. The Calibration resistor value is a 32-bit value
and requires two 16-bit registers. These registers are non-volatile.

Address

Base + 180

16

Bit

15

14

13

12

11

10

09

08

07

06

05

04

03

02

01

00

READ

2nd Least Significant Byte

Least Significant Byte

WRITE

Address

Base + 182

16

Bit

15

14

13

12

11

10

09

08

07

06

05

04

03

02

01

00

READ

Most Significant Byte

2nd Most Significant Byte

WRITE

Value: The 32-bit resistor value (in ohms) is written to two adjacent 16-bit
registers. The value is a float 32 format.

Voltage Calibration

Status Register

The Voltage Calibration Status Register provides a non-volatile record of
the voltage calibration status for each channel. This register is non-volatile.

Address

Base + 184

16

Bit

15

14

13

12

11

10

09

08

07

06

05

04

03

02

01

00

READ

Ch

16

Volt

Cal

Stat

Ch

15

Volt

Cal

Stat

Ch

14

Volt

Cal

Stat

Ch

13

Volt

Cal

Stat

Ch

12

Volt

Cal

Stat

Ch

11

Volt

Cal

Stat

Ch

10

Volt

Cal

Stat

Ch

09

Volt

Cal

Stat

Ch

08

Volt

Cal

Stat

Ch

07

Volt

Cal

Stat

Ch

06

Volt

Cal

Stat

Ch

05

Volt

Cal

Stat

Ch

04

Volt

Cal

Stat

Ch

03

Volt

Cal

Stat

Ch

02

Volt

Cal

Stat

Ch

01

Volt

Cal

Stat

WRITE

Ch XX Cal Stat: This field is set to a one (1) when the channel has been
successfully voltage calibrated. This field is set to a zero (0) when the
channel is not voltage calibrated.

Notes

This register is not set or used by the hardware. The HP E1418A SCPI
driver, if used to calibrate the module, automatically sets the bits in this
register. If you use register programming to calibrate this module, you may
wish to use this register to record the calibration status.

The Voltage Calibration Status Register, the Current Calibration Status
Register, the Calibration Isolation Status Register, and the Calibration Card
Configuration Register are used together to store the overall calibration
status of the module.

144 HP E1418A Register-Based Programming

Appendix B

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Advertising