Calibration – HP E1418A User Manual

Page 76

Advertising
background image

This query returns one of the following values:

Return

Value

Meaning

0

Calibration is complete.

New constants are in RAM.

1

More measurements are needed.

5

Calibration aborted.

A return value of 5 (calibration aborted) is given if either:

a. The command is received when the module is in current calibration or

resistance calibration mode, or

b. The calibration in process has entered more than 50 values without a

successful calibration.

When the channel voltage calibration is completed, the new calibration constants
are used by the module, but not stored in non-volatile memory.

If a voltage calibration is interrupted before completion (that is, by another
command being received), no constants are changed.

Refer to Appendix D, “ Voltage/Current Output Adjustment” , starting on page 154,
for additional information about this command.

Related Commands:

CALibration:STORe

,

CALibration

n

:CONFigure:VOLTage?

CALibration

Chapter 3

HP E1418A SCPI Command Reference 75