Calibration n :value:current, Calibration – HP E1418A User Manual

Page 74

Advertising
background image

CALibration

n

:VALue:CURRent?

CALibration

n

:VALue:CURRent? <

value

>

enters a measured calibration value

and returns a value indicating the present state of the current calibration process.

Parameters

Parameter

Name

Parameter

Type

Range of

Values

Default

CALibration

n

keyword

1 to 8 or 1 to 16

1

<

value

>

numeric (float 32)

0.0 to 6.0 V

none

Returns: unsigned 16-bit integer (uint16).

Comments

In the keyword,

CALibration

n, the valid range of n is:

1 to 8 for

E1418A, E1418A Option 002

1 to 16 for

E1418A Option 001, E1418A Option 003.

Values of n outside these ranges will generate an error.

This command is used together with the

CALibration

n

:CONFigure:CURRent?

query to perform voltage mode calibration. Following the

CAL

n

:CONF:CURR?

query, this command is used repetitively until the return value indicates that the
calibration is complete.

The

<

value

>

parameter is the measured output voltage. A minimum of

9 measurements must be taken to calibrate a channel for current mode output
(a maximum of 50 measurements may be required).

This query returns one of the following values:

Return

Value

Meaning

0

Calibration is complete.

New constants are in RAM.

Resistance is calibrated.

1

More measurements are needed.

Resistance is calibrated.

2

Calibration is complete.

New constants are in RAM.

Resistance is not calibrated.

3

More Measurements are needed.

Resistance is not calibrated.

5

Calibration aborted.

CALibration

Chapter 3

HP E1418A SCPI Command Reference 73