SCPI Command Format

SCPI commands perform functions like setting outputs, querying instrument states,
or retrieving data. The command structure is a hierarchical structure that usually
consists of a top level (or root) command, one or more lower level commands, and
their parameters. The following example shows part of a typical subsystem:

SOURce

n

:

FUNCtion:

MODE <CURRent | VOLTage>

SOURce

n: is the root command (where n is replaced by the channel number in the

range of 1 to 16),

FUNCtion:

is a second level command,

MODE

is a third level

command, and

<CURRent | VOLTage>

is an input parameter.

Command

Separator

A colon (:) always separates one command from the next lower level command.
This is illustrated as follows:

SOURce

n

:FUNCtion:MODE VOLT

Colons separate the root command from the second level (

SOURce

n

:FUNCtion

) and

the second from third level (

FUNCtion:MODE

), and so forth.

Abbreviated and

Short Commands

The command syntax shows most commands as a mix of upper and lower case
letters. The upper case letters indicate an abbreviated spelling for the command.
For shorter program lines, send only the abbreviated form. For better program
readability use the entire command. The instrument will accept either the
abbreviated form or the entire command.

For example, if the command reference syntax shows the command

TRIGger

, then

TRIG

and

TRIGGER

are both acceptable forms. Other forms of

TRIGger

, such as

TRIGG

or

TRIGGE

will generate an error.

The instrument does not distinguish between upper case and lower case characters.
Therefore

TRIGGER

,

trigger

, and

TrigGeR

are all acceptable.

Command keywords can be entered in their full form, as shown above, or can be
entered in their short form. In this manual, the entry required in short form
commands is always capitalized. The short form is generally used for examples in
this manual.