Calibration:condition, Calibration – HP E1418A User Manual

Page 68

Advertising
background image

CALibration:CONDition?

CALibration:CONDition?

queries the module type and calibration status. The

query returns four 16-bit unsigned integers.

Parameters

Returns: four unsigned 16-bit integers (uint16).

Comments

The integers returned indicate the module configuration and status at the last
calibration, not the current configuration of the module. The first and second
integers returned from this query can be compared to the first and third integers
returned from the

DIAGnostic:CONFigure?

query to determine if the module

configuration has changed since the last calibration.

Integer

Returned

Meaning

1

Expansion Board ID in the form:
0 = present, 7 = none
(expansion board contains channels 9–16)

2

2nd 16-bit value indicates how the channels are configured (isolated
or non-isolated). A bit set to 0 indicates an isolated channel, a bit set
to 1 indicates a non-isolated channel. Bits 0–15 correspond to
channels 1–16, respectively.

3

3rd 16-bit value indicates the current calibration status for each
channel. A bit set to 0 indicates the channel is not calibrated for
current, a bit set to 1 indicates the channel has been calibrated for
current. Bits 0–15 correspond to channels 1–16, respectively.

4

4th 16-bit value indicates the voltage calibration status for each
channel. A bit set to 0 indicates the channel is not calibrated for
voltage, a bit set to 1 indicates the channel has been calibrated for
voltage. Bits 0–15 correspond to channels 1–16, respectively.

CALibration

Chapter 3

HP E1418A SCPI Command Reference 67