Conductivity/Resistivity Analyzer/Controller

9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual

7/99

8-2

CELL CAL FACTOR
CAL TRIM C1
CAL TIRM C2
REMOVE CAL TRIM C1
REMOVE CAL TRIM C2

If the Cell 1 solution temperature compensation is set up without TDS, and Cell 2 with TDS,
then the calibration menu will be:

CELL CAL FACTOR
CAL TRIM C1
TDS CVSN FACTOR C2
REMOVE CAL TRIM C1
REMOVE CAL TRIM C2

If both cells’ solution temperature compensation is set up with TDS, then the calibration menu
will be:

CELL CAL FACTOR
TDS CVSN FACTOR C1
TDS CVSN FACTOR C2
REMOVE CAL TRIM C1
REMOVE CAL TRIM C2

Therefore, the screens on your unit may not match exactly the calibration menus used in the
procedure tables in this section.