7 setting the clock – Honeywell 9782 Series User Manual

Page 73

Advertising
background image

I/O Setup and System Configuration

7/99

9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual

5-17

5.7 Setting the Clock

Purpose

A display showing the current time can be accessed when the unit is online. The 24-hour clock
must be set before you can use the AutoClean feature described in Sub-Section 5.6.

Procedure

Follow the procedure in Table 5-7 to set the 24-hour clock.

Table 5-7 Procedure for Setting the Clock

Step

Screen

Action

1

MENU
CONFIGURATION
CALIBRATION
MAINTENANCE
I/O SETUP

DIAGNOSTICS
| PREV | NEXT

Go to CONFIGURATION | ADVANCED FEATURES |
CLOCK SETUP.

CONFIGURATION
CAT/RETRANSMISSION
ALARMS
DISCRETE CONTROL

ADVANCED FEATURES

DIAGNOSTICS
| PREV | NEXT

ADVANCED FEATURES

CLOCK SETUP

COMPUTED VALUE
AUTO CLEAN

DIAGNOSTICS
| |

2

24 HOUR CLOCK SETUP

CLOCK OFF
HOURS 13
MINUTES 30

DIAGNOSTICS
| | NEXT

Set the clock. For example, if you will start the clock at
1:30 PM, enter 13 hours, 30 minutes.

Start the clock and specify what should happen to the
clock if the power fails.

ON W/RESET starts the clock now. Following power

failure, the clock will be reset to 00:00 and the diagnostic
message “RESET CLOCK” will be displayed.

ON W/O RESET starts the clock now. Following power

failure the clock will remain off, and the diagnostic
message “RESET CLOCK” will be displayed.
Display of ”RESET CLOCK” will continue until you either
reset the clock or turn it off.