9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual

10-7

10.4 Viewing Product Information and Changing Model Number
Stored in Memory

Introduction

During I/O setup (see Section 5) the software will display prompts only for those features the
hardware can support. To do this, the software uses the model number of the unit stored in
memory at the factory. Each combination of features has a unique model number (see Section 2).
Therefore, if you add an option card to the 9782, it is very important that you change the model
number stored in memory using the procedure described below.

ATTENTION

A special password is required to change the stored model number. If the password was not
supplied with the new option card, call the Honeywell Technical Assistance Center.

Before calling for technical support, follow the procedure in Table 10-4 to determine the version
of the software in the Analyzer/Controller. The support team will ask for this information.

Procedure

Use the procedure in Table 10-4 to see the software version and model number, and to change
the model number if required.

Table 10-3 Procedure for Viewing Product Information and

Changing the Stored Model Number

Step

Screen

Action

1

MENU
CONFIGURATION
CALIBRATION

MAINTENANCE

I/O SETUP

DIAGNOSTICS
| PREV | NEXT

Go to MAINTENANCE | INSTRUMENT SETUP |
PRODUCT INFO
.

MAINTENANCE
OFFLINE FUNCTIONS

INSTRUMENT SETUP

LEAD RESISTANCE COMP

DIAGNOSTICS
| PREV | NEXT

INSTRUMENT SETUP
INSTURMENT TAGGING

PRODUCT INFO

SCREEN CONTRAST
SECURITY
RESET UNIT
DIAGNOSTICS
| |