Conductivity/Resistivity Analyzer/Controller

9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual

7/99

5-24

PARAMETER

CHOICES or RANGE

(Default is underlined)

I/O setup: discrete control enabled

NO
YES

I/O setup: discrete control type for each available relay

PFT
DAT
O/F [on/off]
N/A

I/O setup: relay activation on alarm/control

ENERGIZE
DE-ENERGIZE

I/O setup: display unit of measure (conductivity only)

standard range models with cell constant

1:

uS (microSiemens per centimeter)
uSm (microSiemens per meter)

standard range models with cell constant of 10,25,or 50:
uS (microSiemens per centimeter) only

wide range models: automatially change from uS to mS
depending on cell constant; no selection available.

I/O setup: relay alarm assignments

ALM1 [alarm 1] default for Relay 1
ALM2 [alarm 2] default for Relay 2
ALM3 [alarm 3] default for Relay 3
ALM4 [alarm 4] default for Relay 4
A1/D [alarm 1 or diagnostic]
A2/D [alarm 2 or diagnostic]
A3/D [alarm 3 or diagnostic]
A4/D [alarm 4 or diagnostic]
A1/2 [alarm 1 or 2]
A3/4 [alarm 3 or 4]
12/D [alarm 1 or 2 or diagnostic]
AA/D [any alarm or diagnostic]
234D [alarm 2 or 3 or 4 or diagnostic]
RNGD [range switch or diagnostic]
AA [any alarm]
RNGS [range switch]

I/O setup: analog output basis [unless DE is enabled]

C1 [Cell 1 value]
C2 [Cell 2 value]
TMP1 [temperature value from Cell 1]
TMP2 [temperature value from Cell 2]

RATO [computed ratio:

Cell

Cell

1

2

]

%REJ [computed percent rejection:

(

)x

1

1

2

100

Cell

Cell

]

%PSG [computed percent passage:

Cell

Cell

x

1

2

100

]

DIFF [computed difference:

Cell

Cell

1

2

]

CO2 [computed parts per million carbon dioxide -
standard range analyzers only]

configuration: retransmission output scaling

LINFIXEDRNG [linear fixed range]

always selected for

computed value
LINAUTORNG [linear auto range]
2DECADELOG
BI-LINEAR