Maintenance

7/99

9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual

10-5

Step

Screen

Action

8

OUTPUT1 SPAN
OUTPUT1 ZERO
OUTPUT2 SPAN
OUTPUT2 ZERO
OUTPUT3 SPAN
OUTPUT3 ZERO
DIAGNOSTICS
| |

The first display will reflect the quantity of analog outputs
supported by the hardware.

Select the value to be calibrated.

Go to the next screen.

9

SPAN CAL OUTPUT1

USE UP/DOWN ARROW
KEYS TO ADJUST
ENTER TO SAVE

DIAGNOSTICS
| |

Use the UP and DOWN keys to correct the reading on the test
meter.

To save the correction press ENTER.

10

OUTPUT CALIBRATION

SAVED

MENU TO EXIT

DIAGNOSTICS
| |

This screen confirms that the value was saved.

11

To calibrate additional values, repeat Steps 1 through 10,
including powering down the unit before changing the
connections to the output terminals.

12

When output calibration has been completed, re-install field
wiring removed in Step 4. Disconnect power before opening
case.

13

Replace the safety cover and secure it with the screw.

14

Close the case and power up the unit. Do not apply power
until the case is closed.