10.3 Specifying a Tagname or Other Display String

Introduction

The real-time displays of process values show the instrument’s tagname (or other configurable
fixed nineteen-character string) at the top of the screen.

Procedure

Use the procedure in Table 10-2 to specify the text to be displayed on the alarm stripe when no
alarms or other messages are active.

Table 10-2 Procedure for Specifying a Tagname or Other Display String

Step

Screen

Action

1

MENU
CONFIGURATION
CALIBRATION
MAINTENANCE
I/O SETUP

DIAGNOSTICS
| PREV | NEXT

Go to MAINTENANCE | INSTRUMENT SETUP |
INSTRUMENT TAGGING
.

MAINTENANCE
OFFLINE FUNCTIONS

INSTRUMENT SETUP

LEAD RESISTANCE COMP

DIAGNOSTICS
| PREV | NEXT

INSTRUMENT SETUP

INSTRUMENT TAGGING

PRODUCT INFO
SCREEN CONTRAST
SECURITY
RESET UNIT
DIAGNOSTICS
| PREV | NEXT

2

UPDATE TAG DATA

HONEYWELL INC

DIAGNOSTICS
| <-- | -->

The text string currently stored in memory will be
displayed.

If desired, edit the string (nineteen characters maximum)
using the UP and DOWN keys, and the <-- and --> function
keys.

Press ENTER or MENU to exit.