4 on-line diagnostics and system error messages – Honeywell 9782 Series User Manual

Page 112

Advertising
background image

Conductivity/Resistivity Analyzer/Controller

9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual

7/99

9-4

9.4 On-Line Diagnostics and System Error Messages

Self-tests at 0.5 second intervals

On-line diagnostics are self-tests which are performed automatically as a background task every
1/2 second. The system error messages (displayed on the alarm stripe if a test is failed) are listed
in Table 9-3.

Output affected by error

In an error condition, the Analyzer will continue to perform all the functions it is capable of
except that the output signal(s) will go to 0%. (Output signal manipulation does not occur with
calibration failure or clock reset diagnostics.)

Table 9-3 On-Line Diagnostics and System Error Messages

SYSTEM ERROR

MESSAGE

WHEN

TESTED

CONDITION

CORRECTIVE

ACTION

A/D CONV FAILURE

background

faulty analog-to-digital converter

Have instrument serviced.

CELL 1 OUT OF RANGE

or

CELL 2 OUT OF RANGE

background

Cell reading is out of range for
either the high or low limit of the
selected cell constant. The
cell’s display will stop updating
until the condition and alarm
message clears. Instead, to
indicate an invalid display, the
display will alternately flash 0.0
and either the last known value
or the high end limit for the cell
constant.

Check cell to make sure it is
connected to terminals. Make
sure cell constant is correct for
expected conductivity range.

CELL 1 TEMP ERROR

or

CELL 2 TEMP ERROR

background

Cell temperature compensator
faulty

Check cell (see Note 1).

CLOCK RESET

after power
failure

Clock is on, but time may be
incorrect, or
clock is off as a result of the
power failure.

Reset the time,
or
if clock is off, turn it on and reset
the time.

EPROM ERROR

at reset

faulty EPROM

Have instrument serviced.

RAM DATA CORRUPTED

background

data in non-volatile memory has
been changed by a fault

Power down, check power supply
and wiring for high voltage
spikes. Have instrument
serviced, if necessary.

RAM ERROR

at reset

faulty RAM

Have instrument serviced.

RAM ERROR IN CPU

at reset

faulty RAM in microprocessor

Have instrument serviced.

Note 1: Temperature compensator should be 8550 ohms at 25

°

C. If temperature compensator

fails, a conductivity instrument will be forced to its display range high limit. A resistivity
instrument will be forced to the low end of the range or zero.