Conductivity/Resistivity Analyzer/Controller

9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual

7/99

10-10

10.6 Entering a Password for Security

Introduction

I/O setup, configuration, calibration and maintenance functions can be password-protected. The
password can be any number between 1 and 9999. (When the password is zero, the operator will
not be prompted to enter a password.)

ATTENTION

The configurable password entered here is not the same as the special password needed to edit
the model number stored in memory as described in 10.5.

Procedure

Follow the procedure in Table 10-5 to enter a password.

Table 10-5 Procedure for Entering a Password

Step

Screen

Action

1

MENU
CONFIGURATION
CALIBRATION

MAINTENANCE

I/O SETUP

DIAGNOSTICS
| PREV | NEXT

Go to MAINTENANCE | INSTRUMENT SETUP |
SECURITY
.

MAINTENANCE
OFFLINE FUNCTIONS

INSTRUMENT SETUP

LEAD RESISTANCE COMP

DIAGNOSTICS
| PREV | NEXT

INSTRUMENT SETUP
INSTURMENT TAGGING
PRODUCT INFO
SCREEN CONTRAST

SECURITY

RESET UNIT
DIAGNOSTICS
| |

2

PASSWORD SECURITY

ENABLE SECURITY: YES
PASSWORD: 0000

DIAGNOSTICS
| |

To enable the security function, select YES.

The current password will be displayed. It can be edited
using this screen.