Specifications and Model Number

7/99

9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual

2-3

Wide range models

Wide Display Ranges: 9782C-W0

Conductivity (microsiemens/cm)

TDS

Cell Constant Required (cm-1)

100-1,999

0-1,999 ppm

0.1

1000-19,990

0-19,990 ppm

1

Conductivity (millisiemens/cm)

10-199.9

0-199.9 ppt

10

10-500

25

20-1,000

50

Concentration

0-18.00% Hydrochloric Acid*

50**

0-19.99% Sulfuric Acid

50**

0-19.99% Sodium Hydroxide*

25** ro 50

0-19.99% Sodium Chloride

25** or 50

* The maximum concentration measurable by conductivity depends on temperature. Maximum

for sodium hydroxide allowed by the 9782 Analyzer is 10% at 0º C, 13% at 25º C and 20% at
75º C with linear interpolation between these points. The maximum concentration for
hydrochloric acid is 15.5% above 40º C.

** Measurements in the lower half of the concentration ranges and below 50º C may also use

the next lower available cell constant.

2.1.6 Operating Conditions

Operating Conditions

Ambient Temperature: Normal 0 to 60

°

C (32 to 140

°

F), extreme -10 to +60

°

C (14 to 140

°

F);

storage -30 to +70

°

C (-22 to 158

°

F).

Line Voltage: Normal 120 +/- 10% Vac (extreme 100 to 132 Vac, 47 to 63 Hz)
Normal 240 +/- 10% Vac (extreme 200 to 264 Vac, 50 to 60 Hz)

RH: 90% maximum non-condensing @ 40

°

C (104

°

F) max.

Installation Category: II

Pollution Degree: 2

Altitude: <20,000 M (6562 ft)