9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual

2-5

2.1.8 Temperature Compensation

Availability

All models can provide high resolution (0.05º C, typical) temperature compensation by
microprocessor calculation over 0 to 140º C for high purity water and concentrations as well as
routine water measurements.

standard range models: Temperature compensation algorithms are for NaCl,
acid/cation/ammonia, or morpholine.

wide range models: Temperature compensation algorithms are for NaCl, HCl,
NaOH, or H

2

SO

4

.

Displayed process value can be toggled between compensated and uncompensated while the
Analyzer/Controller is online.

2.1.9 AutoClean

Cell rinse schedule settings

Timer Cycle: Adjsutable; up to 28 days with day, hour, minute resolution.
Rinse Time: 1-1999 seconds.
Resume Process Time Delay: 1-1999 seconds.

2.1.10 Control

Settings

On/off period: 1 to 1000 seconds
On/off percent “on” time: 0 to 100%, 1% resolution.
Setpoint and proportional band limit ranges:conductivity: 0.055 to 19,999

µ

S/cm

resistivity: 0 to 19.99 Mohms/cm
temperature: 10 to 140

°

C

DAT cycle period: 1 to 200 seconds.
PFT maximum frequency: 1 to 200 pulses/minute.
PFT pulse width: 50 ms, compatible with electronic pulse-type metering pumps.