Conductivity/Resistivity Analyzer/Controller

9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual

7/99

5-26

PARAMETER

CHOICES or RANGE

(Default is underlined)

configuration: AutoClean current day

0 to 28, default = 0

configuration: AutoClean rinse duration
configuration: AutoClean resume delay

1 to 1999 seconds, default = 60

configuration: computed value selection

NONE

RATIO [computed ratio:

Cell

Cell

1

2

]

DIFFERENCE [computed difference:

Cell

Cell

1

2

]

%PASSAGE [computed percent passage:

Cell

Cell

x

1

2

100

]

%REJECTION [computed percent rejection

:

(

)x

1

1

2

100

Cell

Cell

]

PPMCO2 [computed parts per million carbon dioxide -
9782C-S0 only]

maintenance: display resolution

LOW
HIGH

maintenance: security enabled

YES
NO

maintenance: password

0000 to 9999; default = 0000

maintenance: lead resistance compensation wire gauge
(AWG)
wide range models only

12
14
16
18

maintenance: lead resistance compensation length
wide range models only

1 to 1500 feet default = 1

calibration: TDS conversion factor

standard range models with cell constant of 1, 10, 25, or
50:
0.01 to 1.999; default = 0.5 ppm
standard range models with cell constant of 0.01 or 0.1:
10 to 1999; default = 500 ppb

wide range models for all cell constants:
0.01 to 1.999; default = 0.5 ppm