General calibration – Garland 200 User Manual

Page 40

Advertising
background image

Page 40

Part # MCOSM06 Rev 1 (11/03/08)

GENERAL CALIBRATION

Instrumentation

A thermocouple type test instrument is preferred for

measuring oven temperatures accurately. Mercury

thermometers are acceptable providing they can be proven

accurate.

Regardless of the type instrument used, it is most desirable

to double-check it just before making an oven temperature

check. This can be done simply and quickly by placing

the thermocouple tip (or immersing the entire mercury

thermometer) in boiling water. Note: Mercury oven

thermometers should be the "Total immersion type." The

resulting reading should be within several degrees of 212°F,

depending on the altitude.

Generally, a mercury thermometer can drop and excess of

25° F when the doors are opened to check oven temperature.

This condition makes it extremely difficult to get a true

temperature reading.

Almost all convection ovens today have oven burners which

cycle "on and off" The high and low points of the resulting

temperature "swing" (differential) in the oven must be

measured- and averaged-to determine the true operating

temperature. A thermocouple type test instrument is best

suited to measure these temperature changes quickly and

accurately, and without opening the oven door.

With an accurate, thermocouple test instrument or mercury

thermometer, it is still difficult to measure these changing

oven temperatures and then, average correctly. This is why

we recommend that the thermocouple tip or the bulb of the

thermometer be "weighed" Weighing (adding mass) to the

test instrument compensates for oven temperature changes

making the test instrument less sensitive to these constant

changes in temperature. Note: How this weighting can be

done is discussed later.

Measuring these changing high and low temperature points

in an oven is possible with either type of test device with out

weighing, but is subject to possible inaccuracies. This is most

true at a low temperature setting because, in this area, the

function of "time" becomes a factor. The simple averaging

of temperatures then may not produce the true operating

temperature. "Weighing" provides the compensation for both

time and temperature that is necessary. We produced test

curved showing actual results in the low temperature area

showing the difference in results when using an unweighted

and weighted thermocouple or thermometer. Tests were

made with the same control, same oven and dial set at 170°

mark (not moved).

260

240

220

200

180

160

140

120

100

Chart No.1

Unweighted Thermocouple dial setting 170 F

260
240
220
200
180
160
140
120
100

Chart No.2

Weighted Thermocouple Dial Setting 170F

From the above, it can be seen that an error of 15°F is

possible. Chart #1, unweighted thermocouple indicates

an erroneous average oven temperature of 185°F, Chart

#2 weighted thermocouple indicates the "average" or true

temperature to be 170°F.

Weighing of Thermocouple

A thermocouple can be weighted by clamping the tip

between two one inch square 1/16" inch thick aluminum. The

thermocouple can also be weighted (just as successfully) by

using a letter sized sheet of aluminum foil. Fold the foil five

times doubling the thickness each fold. After the fifth fold,

place the thermocouple tip in the center of the aluminum

piece and fold once more. Finally fold in the sides so that the

foil clings to the thermocouple tip. A mercury thermometer

can be weighted in much the same way by wrapping several

layers of aluminum foil around the bulb end thus creating the

necessary mass. This procedure is a must if you open the door

to check the temperature.