8.1.2 Measuring the temperature

For reproducible conductivity measurements, it is essential to measure the
temperature of the test sample.
IDS sensors measure the temperature with a temperature sensor integrated in
the IDS sensor.

8.2

Temperature compensation

The calculation of the temperature compensation is based on the preset refer-
ence temperature, 20 °C or 25 °C. It appears on the display as Tr20 or Tr25.
You can select one of the following temperature compensation methods:

Nonlinear temperature compensation (nLF) according to EN 27 888

Linear temperature compensation (Lin) with adjustable coefficients of
0.000 ... 3.000 %

/

K

No temperature compensation (off)

Application tips

Select the following temperature compensations given in the table according to
the respective test sample:

The reference temperature and temperature compensation are set
in the menu for the parameter, conductivity (see section 9.4.1 S

ET

-

TINGS

FOR

IDS

CONDUCTIVITY

SENSORS

, page 66).

Test sample

Temperature compensation

Display

Natural water
(ground water, surface
water, drinking water)

nLF
according to EN 27 888

nLF

Ultrapure water

nLF
according to EN 27 888

nLF

Other aqueous solu-
tions

Lin
Set linear temperature coefficient
0.000 ... 10.000 %/K

Lin

Salinity (seawater)

Automatic nLF according to IOT
(International Oceanographic
Tables)

Sal, nLF