4 cond measurement settings, 1 settings for ids conductivity sensors, Cond measurement settings – YSI MultiLab 4010P-1 User Manual

Page 66: Settings for ids conductivity sensors, Ctivity (see section 9.4.1 s, Tings, Conductivity, Sensors

Advertising
background image

Settings

MultiLab 4010-1

66

ba76141e03

03/2014

9.4

Cond measurement settings

9.4.1 Settings for IDS conductivity sensors

Settings

The settings are made in the menu for the measured parameter, conductivity.
To open the settings, display the required measured parameter in the mea-
sured value display and press the <MENU/ENTER> key. After completing the
settings, switch to the measured value display with <M>.
The possible settings are individually displayed for each sensor. Below the set-
ting menu is displayed for two IDS sensors (

4310, 4320).

Default settings are printed in bold.

Setting menu of

4310

Menu item

Possible
setting

Description

Calibration /Cali-
bration record

-

Displays the calibration record of the
last calibration.

Calibration / Cali-
bration data stor-
age
/ Display

-

Displays the last calibration records
(max. 10)

Calibration /
Calibration data
storage
/ Output to
RS232/USB

-

Outputs the calibration data storage to
the interface

Calibration /Cali-
bration interval

1 ... 150 ...
999 d

Calibration interval for the IDS con-
ductivity sensor (in days).
The meter reminds you to calibrate
regularly by the flashing sensor sym-
bol in the measuring screen.

Type

Measuring cell used

cal

Measuring cells whose cell constant is
determined by calibration in the KCL
control standard solution.
Calibration range:
0.450 to 0.500 cm

-1

The currently valid cell constant is dis-
played in the status line.

man

Manually adjustable cell constant in
the range 0.450 ... 0.500 cm

-1

.

Cell const. man

0.450 ...
0.475 ...
0.500 cm

-1

Display and setting options for the
manually adjustable cell constant.

Temp. comp. (TC) /
Method

nLF
Lin
off

Procedure for temperature compen-
sation (see section 8.2 T

EMPERATURE

COMPENSATION

, page 55).

This setting is only available for the
measured parameters, conductivity
(

ϰ

) and resistivity (

ρ)

.