9 settings, 1 ph measurement settings, 1 settings for ph measurements – YSI MultiLab 4010P-1 User Manual

Page 59: Settings, Ph measurement settings, Settings for ph measurements, Selected (see section 9.1.1 s, Ettings, Measurements, See section 9.1.1 s

Advertising
background image

MultiLab 4010-1

Settings

ba76141e03

03/2014

59

9

Settings

9.1

pH measurement settings

9.1.1 Settings for pH measurements

Settings

The settings are made in the menu for calibration and measurement settings of
the pH/ORP measurement. To open the settings, display the required mea-
sured parameter in the measured value display and press the <MENU/
ENTER>
key. After completing the settings, switch to the measured value dis-
play with <M>.
Default settings are printed in bold.

Menu item

Possible
setting

Description

Calibration / Calibra-
tion record

-

Displays the calibration record of the last calibration.

Calibration / Calibra-
tion data storage
/ Dis-
play

-

Displays the last calibration records (max. 10)

Calibration /
Calibration data stor-
age
/ Output to
RS232/USB

-

Outputs the calibration data storage to the interface

Calibration / Buffer

YSI
ConCal
NIST/DIN
...

Buffer sets to be used for pH calibration.
More buffers and details: see section 9.1.2 B

UFFER

SETS

FOR

CALIBRATION

, page 60 and section 5.2

P

H

CALIBRATION

, page

30.

Calibration / Single-
point calibration

yes
no

Quick calibration with 1 buffer

Calibration / Calibra-
tion interval

1 ... 7 ...
999 d

Calibration interval for the IDS pH sensor (in days).
The meter reminds you to calibrate regularly by the flashing
sensor symbol in the measuring screen.

Calibration / Unit for
slope

mV/pH
%

Unit of the slope.
The % display refers to the Nernst slope of
-59.2 mV/pH (100 x determined slope/Nernst slope).

QSC / First calibration

-

Starts the initial calibration with QSC buffers.
This menu item is only available as long as no initial calibra-
tion was carried out with the connected IDS sensor.

QSC / Record of first
calibration

-

Displays the calibration record of the QSC initial calibration.

QSC / Control calibra-
tion

-

Starts the control calibration with QSC buffers.
This menu item is only available if an initial calibration was
carried out with the connected IDS sensor.

Man. temperature

-25 ... +25
... +130 °C

Entry of the manually determined temperature.
This menu item is only available if an IDS adapter is con-
nected.