2 buffer sets for calibration, Buffer sets for calibration – YSI MultiLab 4010P-1 User Manual

Page 60

Advertising
background image

Settings

MultiLab 4010-1

60

ba76141e03

03/2014

9.1.2 Buffer sets for calibration

You can use the buffer sets quoted in the table for an automatic calibration. The
pH values are valid for the specified temperature values. The temperature
dependence of the pH values is taken into consideration during the calibration.

pH resolution

0.001
0.01
0.1

Resolution of the pH display

mV resolution

0.1
1

Resolution of the mV display

Reset

-

Resets all sensor settings to the delivery condition (see sec-
tion 9.6.1 R

ESETTING

THE

MEASUREMENT

SETTINGS

, page 70).

Menu item

Possible
setting

Description

No.

Buffer set *

pH values

at

1

YSI *

4.000
7.000
10.000

25 °C

2

ConCal

Any

Any

3

NIST/DIN
DIN buffers according to DIN 19266
and NIST Traceable Buffers

1.679
4.006
6.865
9.180
12.454

25 °C

4

TEC
Technical buffers

2.000
4.010
7.000
10.011

25 °C

5

Merck 1*

4.000
7.000
9.000

20°C

6

Merck 2 *

1.000
6.000
8.000
13.000

20°C

7

Merck 3 *

4.660
6.880
9.220

20°C

8

Merck 4 *

2.000
4.000
7.000
10.000

20°C