9.1.3 Calibration interval

The calibration evaluation is displayed as a sensor symbol.

After the QSC function has been enabled the sensor symbol is replaced by the
QSC scale (see section 5.2.8 QSC

FUNCTION

(

SENSOR

QUALITY

CONTROL

), page

40).

After the specified calibration interval has expired the sensor symbol or the
QSC scale flashes. It is still possible to measure.

Setting the

calibration interval

The calibration interval is set to 7 days in the factory.
You can change the interval (1 ... 999 days):

22

Reagecon 20 *

2.000
4.000
7.000
10.000
13.000

20°C

23

Reagecon 25 *

2.000
4.000
7.000
10.000
13.000

25 °C

24

Riedel-de Haen *

2.000
4.000
7.000
10.000

20 °C

25

USABlueBook

*

4.000
7.000
10.000

25 °C

The buffers are selected in the menu, pH / <MENU/ENTER> / Cal-
ibration
/ Buffer (see section 9.1.1 S

ETTINGS

FOR

P

H

MEASURE

-

MENTS

, page 59).

To ensure the high measuring accuracy of the measuring system,
calibrate after the calibration interval has expired.

No.

Buffer set *

pH values

at

1.

Open the menu for measurement settings with <MENU/ENTER>.

2.

In the Calibration / Calibration interval menu, set the calibration interval
with <><>.