12.5 What to do if... / printer

Integrated printer

does not print

Printer operating -

paper not being

printed

Integrated printer

prints automatically

Cause

Remedy

– Printer switched off

(LED flashing)

– Switch on the printer

(LED illuminated)

– No power pack connected

– Connect the power pack

– USB cable connected

– Disconnect the USB cable from the

meter

– Function "automatic storing at

intervals" with long intervals is
switched on

– Switch off the function (see section

10.2 A

UTOMATIC

DATA

STORAGE

AT

INTERVALS

, page 74)

– No paper available

– Insert a roll of paper

Cause

Remedy

– Paper inserted the wrong way up – Turn the roll of paper around and

insert it with the other side upwards

Cause

Remedy

– The function "Automatic storage

at intervals" or "Transmit data
automatic at intervals" is
switched on

– Switch off the function (see section

10.2 A

UTOMATIC

DATA

STORAGE

AT

INTERVALS

, page 74 or

section 11.4 O

PTIONS

FOR

DATA

TRANSMISSION

TO

A

PC, page 80)

Advertising
This manual is related to the following products: