Appendix B SDI-12 Configuration for a Hydrolab Sonde

Prerequisites:

SDI-12 compatible sonde

Power adapter (Mains/Battery. Power adapter not required if
sonde is powered by Internal Battery Pack.)

External power adapter cable (not required if sonde is powered
by Internal Battery Pack).

Detachable cable (Cat. no. 13470HY/015XXX)

HYDRAS 3LT software tool installed in PC

PC

Refer to

Figure 42

for when configuring the power supply setup.

Setup for sonde powered by IBP

If your sonde is internally powered by an IBP (Internal Battery
Pack), connect the 9-pin end of the detachable cable to one of the
COM ports available in the PC. Connect the other end of the
detachable cable (6-pin connector) to the bulkhead connector.
Follow the SDI-12 configuration instructions (see

SDI-12

Configuration on page 104

).

Setup for sonde with external power source

Connect the power adapter to the power source (AC main or
external battery) and connect the other end to the external power
adapter cable. Connect one end of the external power adapter
cable to one of the COM ports in the PC and the other end to the
detachable cable. Connect the 6-pin connector to the bulk head
connector of the sonde. Follow the instructions given for configuring
the SDI-12 (see

SDI-12 Configuration on page 104

).

Figure 42 Setup for sonde with external power source

1

To power source (if applicable)

4

Sonde bulkhead connector

2

9-pin connector

5

6-pin connector

3

Cable configuration may vary depending on application. Refer to Hydrolab DS5X, DS5, and MS5 Water Quality
Multiprobes User Manual (Cat. No. 003078HY).