54

Sampler operation

2. The sample history will list the sample numbers, bottle numbers

and brief result for each sample. Select a sample from the list
and press ENTER to view complete details.

3. The complete details for the selected sample will be shown.

4.7.5 Event log

The event log records information on the following events:

Power on—power was connected to the controller

Power fail

—power was intentionally or unintentionally

disconnected from the controller without first turning power off

Firmware update—a new version of firmware was installed

Pump fault—a fault occurred when the pump was operating

Arm faulty—the distributor arm did not move correctly

Low memory battery—the internal battery should be replaced

User on—a user turned the power on using the power key

User off—a user turned the power off using the power key

Program start—the sample program was started

Program resume—a program was resumed from the halt state

Program halt—a program was halted

Program complete—a program was completed

Grab sample—a grab sample was taken

Change tube—the maximum pump cycle counts have expired

Cooling fail

—the temperature of the thermal unit has been

above the setpoint continuously for more than 30 minutes

Heating fail

—the temperature of the thermal unit has been

below the setpoint continuously for more than 30 minutes

Thermal error clear

—a Cooling Fail or Heating Fail error has

been corrected

SDI-12 timeout:—indicates a communication error on the SDI-12
interface

High setpoint on— Indicates that a user defined high setpoint is
activated, the measurement channel of interest, and the current
measurement value.

High setpoint off— Indicates that a user defined high setpoint is
cleared, the measurement channel of interest, and the current
measurement value.

Low setpoint on— Indicates when a user defined low setpoint is
activated, the measurement channel of interest, and the current
measurement value.

Low setpoint off— Indicates when a user defined low setpoint is
cleared, the measurement channel of interest, and the current
measurement value.