3 rotor removal and cleaning, Maintenance 6.3.3 rotor removal and cleaning – Hach-Lange SIGMA SD900 Refrigerated Sampler and All Weather Refrigerated Sampler User Manual User Manual

Page 81

Advertising
background image

79

Maintenance

6.3.3 Rotor removal and cleaning

For an entire pump assembly replacement, order part number 6262000.

Remove the pump rotor to thoroughly clean the silicone residue
from the pump tube.

Procedure:

1. Remove the pump tube (see

section 6.3.2 on page 76

).

2. Open the rubber flap on the end of the rotor and remove the

screw with a flat-blade screwdriver (

Figure 35

).

3. Remove the rotor. A collet connects the rotor to the pump shaft

and can come off when the rotor is removed (

Figure 35

).

4. Clean the rotor, pump tube tracks and pump housing with a

mild detergent.

5. If the collect was removed in step

3.

, align the recess in the

collet with the tang on the shaft and push the collet on the shaft
(

Figure 35

).

6. Push the rotor on the shaft and secure with the screw. Do not

overtighten the screw. Close the rubber flap on the rotor.

7.

Install the pump tube (

section 6.3.2 on page 76

)

.

Figure 35 Pump rotor removal

1

Shaft

5

Rotor

2

Collet recess

6

Collet

3

Rubber flap

7

Tang

4

Screw

Advertising