65

Section 5

Advanced sampling

5.1 Advanced sampling overview

Use the advanced sampling menu for the following operations:

Send a 12 VDC output signal to start a second sampler,
activate a relay or for communications

Set the sampler to start and stop from an external signal

Set multiple start and stop times

Set variable time intervals

Collect first flush samples from stormwater

Define a setpoint sampling algorithm.

Configure the sampler with timed bottle sets.

5.2 Advanced sampling menu

Table 5

outlines the menu for the advanced sampling operations.

Table 5 SD900 advanced sampling menu

Menu option

Description

Program complete O/P

Send a signal at the completion of a sampling program. The signal can be used to start a second
sampler or activate a relay.

Setpoint sample

Start and stop the sampler from either an external trigger on the auxiliary interface or a
measurement supplied on the SDI-12 interface.

Special output

Send a signal to a connected device.

Each sample

Send signal at the completion of each sample cycle.

Sample pumping

Send signal during the sample intake.

Rinse to purge

Send signal during the entire sample cycle, including all purge and rinse cycles.

Start/stop times

Start and stop a program at user specified dates and times.

Program start/stop

Set up to 12 start and stop times for any one program.

Stormwater

Collect first flush samples at timed intervals. Can be initiated by external device.

Immediate

Start program immediately.

External

Start program from an external device.

Variable intervals

Sample at variable time or flow intervals.

Timed Bottle

Set the sample collection to timed bottle.

Timed bottle samples

Set the number of samples to be collected to a timed bottle set.

24 hr format

Set the timed bottle trigger time in 24 hour format.

Duration

Set the timed bottle trigger in duration of hour and minutes

Continuous

Sampler will continue collecting samples in timed bottle sets until stopped manually

End after last

The controller stops sampling when the last sample is collected in the last bottle of last timed
bottle set.

Done

Exit the advanced sampling menu.