Maintenance

6.6 Thermal sensor calibration (AWRS only)

6.6.1 Thermal sensor verification

To verify the accuracy of the refrigerator compartment temperature,
which is indicated on the sampler controller display, place a
laboratory grade thermometer into a beaker with 150 mL water into
the right rear corner of the tray, and then close the refrigerator
compartment door.

Allow the refrigerator to run for 6–7 hours. This will make sure the
equilibrium between the temperature in the compartment and the
water in the beaker.

Compare the refrigerator temperature readout on the thermal
control diagnostic screen (select

DIAGNOSTICS>THERMAL

DIAGNOSTICS

from the main menu) with the temperature indicated

on the thermometer. If the readings do not agree, it is necessary to
calibrate the Reference Temperature Sensor. This cylindrical
shaped sensor is located in the refrigerated compartment right rear
corner, and is held in place by a clip bracket.

6.6.2 Thermal sensor calibration

Prerequisites:

Ice bath in a 500 mL beaker, using a 50/50 mix of crushed ice
and water

Lab grade thermometer

Procedure:

1. Press the

POWER

button to shut off the sampler controller.

2. Remove pull out tray.

3. Remove the Reference Temperature Sensor (RTS) from the

bracket, and place the beaker/ice bath next to the bracket.

4. Place the sensor and the lab grade thermometer in the

beaker/ice bath.

Figure 38 Desiccant holder

1

Plug

3

O-ring

2

Desiccant tube