10 fuse listing, Maintenance – Hach-Lange SIGMA SD900 Refrigerated Sampler and All Weather Refrigerated Sampler User Manual User Manual

Page 88

Advertising
background image

86

Maintenance

6.10 Fuse Listing

WARNING
Potential Shock and Fire hazards. Only qualified personnel
should conduct the tasks described in this section of the
manual. Always disconnect power to the sampler before
performing any maintenance or service.

DANGER
For continued protection against fire, replace fuses only with
fuses of the same specified type and current rating.

Fuse information for the sampler is listed in

Table 6

.

Note: The fuses are not serviceable by the operator.

Table 6 Fuse listing

Operator

accessible

Location

Reference

designator

Circuit

protection

Size

Rating

Hach part

number

No

Thermal control

circuit board

(6251400) in

the power

control box

under the rear

cover

(

Figure 39 on

page 84

).

F1

Power supply

output

(15 VDC)

¼ x 1 ¼ inch

T, 7A, 250V

6257700

F2 and F3

AC-DC power

supply (AC)

5 x 20 mm

T, 2A, 250V

6257600

F4 and F5

Optional

controller

heater (AC)

T, 5A, 250V

6610000

Circuit board

(8732800)

inside of SD900

controller

F2

SD900

circuits DC

T, 7A, 125 V

8748800

Advertising