46

Sampler operation

4.4.1 Sampler program overview

Table 3

outlines the menu for creating a basic sampling program.

Table 3 Basic program setup

Menu option

Description

Bottles

Enter bottle information.

Bottle quantity

Select the number of bottles in the sampler (1, 2, 4, 8 or 24).

Bottle volume

Enter the volume capacity of each bottle in milliliters or gallons (0.5–99.8 gal or 50–65000 mL).

Intake tubing

Enter intake tubing information.

Tubing length

Enter the length of the intake tube from

the strainer to the liquid sensor (3–99 ft or

100–3000 cm). An accurate length is necessary to obtain accurate sample volumes.

Intake tube type

Select the size and type of tubing (¼ in. vinyl,

3

/

8

in. vinyl or

3

/

8

in. teflon)

Program delay

Delay starting the sampling program until specified time and day or until specified number of
counts is reached.

Enable/disable

Select enable to use a program delay or disable to omit it.

Date and time

If enabled, enter the date and time when the program will start (24 hour format).

Counts

If enabled, enter the counts from the flow meter when the program will start (1–9999 counts). If
counts is selected and the sample pacing is later set to time, the program delay will be disabled.

Sample pacing/collection

Specify whether to collect samples at regular time intervals or regular flow volume.

Time based

Collect samples at regular time intervals.

Pacing interval

Enter the time interval, in hours and minutes (0:01–999:00).

Take first sample

Select whether the program will start immediately or after the first time interval has passed.

Flow based

Collect samples at a specified flow volume (requires external flow meter).

Take sample every

Enter the flow interval that must expire between sample cycles, in counts (1–9999 counts).

Override time

Select enable to force a sample to be collected if the flow volume is unusually low.

Time

If enabled, enter the maximum time between samples (0:01–999:00). Timer is reset each time a
sample is collected based on flow counts.

Take first sample

Select whether the program will start immediately or after the first flow interval has passed.

Sample distribution

For multiple bottle sampling, specify how the samples are delivered to the bottles.

Deliver samples to all

Select whether each sample will be delivered to all bottles or not.

Yes

Each sample will be delivered to all bottles.

End after last

Program will stop after the last sample has been collected. Enter number of samples (1–999).

Continuous

Sampler will continue collecting samples until stopped manually.

No

Samples will be delivered to a subset of bottles (see

Figure 27 on page 48

).

Samples/bottle

Enter the number of samples to be collected in each bottle (1–999).

Bottles/sample

Enter the number of bottles that will contain the same sample.

Liquid sensor

Enable/disabled. If disabled, the sample volume must be calibrated by time.

Sample volume

Enter the volume to be collected per sample. If bottles/sample mode is selected, each bottle
receives a full sample volume (100–10,000 mL).

Intake rinses

Enter the number of times the intake tube will be rinsed before each sample is collected (0–3).

Sample retries

Enter the number of sample retries to perform when a failure occurs (0–3).

Site ID

Enter a name for the sampling location (enter up to 12 characters). The site ID is used as the
preset name if the program is saved.

Advanced sampling

Access the advanced sampling menu.

Done

Finished creating basic sampling program. Prompts user to start or cancel the program.