72

Advanced sampling

5.3.6 Set variable intervals

Variable time or flow intervals can be set for sample cycles.
Intervals will be in hours and minutes if the sample pacing mode is
time based. Intervals will be in counts if the sample pacing mode is
flow based.

Procedure:

1. Select

PROGRAM SETUP

>

MODIFY SELECTED

>

ADV SAMPLING

from the main menu.

2. Select

VARIABLE INTVL

>

ENABLE

.

3. Enter the time (or count) interval between each sample cycle.

Up to 99 intervals can be entered.

Note: When entering individual intervals the current interval is recorded
when the enter key is pressed.

5.3.7 Timed bottle sets

The timed bottle option allows a single sampler unit to function like
multiple samplers. This can be useful in situations such as high flow
versus low flow days.

In this mode, the user configures the sampler to execute a sampling
program in multiple parts. The first part of the program executes for
a time interval defined by the operator. The second part then
operates for a user defined time interval, followed by the third part,
and so on.

The sample bottles are also divided into multiple sets. The first set
of bottles is assigned to part one of the sampling program. The
second set of bottles is assigned to part two of the sampling
program, and so on.

A timed bottle program can run at the same time as the storm
program, setpoint sampling, and variable intervals.

The timed bottle function, when enabled with variable intervals, will
collect samples in each timed bottle set according to the variable
intervals defined. When a timed bottle set switch occurs upon
expiration of the trigger time, the variable interval index is reset.

Note: Timed bottle can be enabled only if the sampler is configured with
multiple bottles, non-composite mode, and the distribution mode is set to
Bottles/Sample.

Procedure:

1. Select

PROGRAM SETUP>MODIFY SELECTED>ADVANCED

SAMPLING

from the main menu.