4 distributor arm tubing replacement, Maintenance – Hach-Lange SIGMA SD900 Refrigerated Sampler and All Weather Refrigerated Sampler User Manual User Manual

Page 82

Advertising
background image

80

Maintenance

6.4 Distributor arm tubing replacement

The distributor arm moves over each bottle during multiple bottle
sampling. Inspect the tubing in the distributor arm on a regular
basis. Replace the tubing when deteriorated or at regular intervals.

Prerequisites:

Tubing, distributor arm, pre-cut. Make sure that the correct tube
is used for the correct distributor and distributor arm. Refer to

Figure 3 on page 11

or

Figure 4 on page 12

and to

Replacement Parts and Accessories on page 93

.

Procedure:

1. Remove the old tubing from the distributor arm.

2. Insert the new tubing into the distributor arm so that the end of

the tubing extends out of the nozzle no more than

1

/

8

inch

(

Figure 36

).

Note: Do not let the tubing extend more than

1

/

8

inch past the nozzle

end of the arm. If the tubing extends too far, the tubing will get caught
on the bottles and interfere with sampling.

3. To make sure that the distributor is aligned properly, run the

manual distributor diagnostic (see

Diagnostic tests on

page 90

).

Figure 36 Distributor assembly

1

Nozzle

4

Shaft

2

Distributor arm

5

Distributor tube

3

Distributor motor

Advertising