3 operate from external signal (setpoint sample), Advanced sampling – Hach-Lange SIGMA SD900 Refrigerated Sampler and All Weather Refrigerated Sampler User Manual User Manual

Page 69

Advertising
background image

67

Advanced sampling

During sample intake—during the sample intake portion of the
cycle only, ignoring all purges and rinse cycles.

During entire cycle—during the entire sample cycle including all
purge and rinse cycles.

Procedure:

1. Select

PROGRAM SETUP

>

MODIFY SELECTED

>

ADV SAMPLING

from the main menu.

2. Select

SPECIAL OUTPUT

>

ENABLE

.

3. Select one of the output options:

AFTER CYCLE

—at the completion of each sample cycle.

SAMPLE PUMPING

—during sample intake only.

ENTIRE CYCLE

—during entire sample cycle.

5.3.3 Operate from external signal (setpoint sample)

The setpoint sample option starts and stops the sampler after
receiving a signal from an external device such as a flow meter.
The current status indicates when the sample program is being
inhibited by an external device.

Note: Setpoint and stormwater sampling cannot be enabled at the
same time.

Procedure:

1. Select

PROGRAM SETUP

>

MODIFY SELECTED

>

ADV SAMPLING

from the main menu.

2. Select

SETPT SAMPLE

>

SETPT TRIGGER>EXT TRIGGER

>ENABLE

.

3. Select SETPT SAMPLE>SETPT CONTROL and select one of

the setpoint options:

START ON SETPT

starts a sample program when the

signal is received, ignoring all future transitions, and
continues until the program is complete.

START/STOP

starts a sample program when the signal is

active and stops the program when the signal is no longer
active.

Note: SETPT CONTROL cannot be set unless a setpoint trigger is
enabled.

4. Enter a delay time in hours and/or minutes that must expire

before the sample program starts. The delay time will begin
when the external signal is received. The program will begin
after the delay time has passed.